Diagnostica Stago STA Compact Max Bi-Directional LIS Interfacing Guide

Complete engineering specification, HL7/ASTM protocols, pinouts, and parameter mapping for Diagnostica Stago STA Compact Max on AILS Cloud LIMS.

Diagnostica Stago STA Compact Max Analyzer - AILS LIS Interfacing Guide
Manufacturer
Diagnostica Stago
Protocol
ASTM E1394 / Serial Data Link
Interface Mode
Bi-Directional
Throughput
Automated Hemostasis Runs

1. Diagnostica Stago STA Compact Max Interfacing Overview & Diagnostic Workflow

The Diagnostica Stago STA Compact Max is a precision diagnostic analyzer designed for clinical laboratory testing. Connecting the Diagnostica Stago STA Compact Max directly to AILS Cloud LIMS eliminates manual data entry, prevents typographical identification errors, and delivers verified results directly to ordering physicians and hospital EMRs.

Bi-Directional Order Interfacing

AILS Gateway manages bi-directional ASTM E1394 / Serial Data Link worklist handshakes with the Diagnostica Stago STA Compact Max, streaming patient test requests as samples are loaded.

Quality Control & Reference Range Interlock

Results are checked against age- and gender-adjusted reference ranges, delta check baselines, and quality control (QC) Levey-Jennings rules before digital sign-off.

2. Diagnostica Stago STA Compact Max RS-232 Serial Port & Pinout Configuration

Hardware & Socket Configuration

RS-232 DB9 Female Serial Cable (COM1 / 9600 Baud) or TCP/IP Network Link.

Protocol: ASTM E1394 / Serial Data Link
Interface Mode: Bi-Directional
Default Speed: 9600 bps, 8-N-1
// RS-232 DB9 NULL-MODEM PINOUT SERIAL COM
[Instrument DB9]                   [AILS Gateway PC]
Pin 2 (TxD / Data Out) ---------> Pin 2 (RxD / Data In)
Pin 3 (RxD / Data In)  <--------- Pin 3 (TxD / Data Out)
Pin 5 (GND / Signal)   <========> Pin 5 (GND / Signal)
Pin 7 (RTS) & Pin 8 (CTS)        [Loopback or None]
Baud: 9600 bps | Data Bits: 8 | Parity: None | Stop: 1
Verified with AILS Gateway Daemon & Real-Time Socket Router.

3. Diagnostica Stago STA Compact Max ASTM E1394 Record Frame

Below is a validated communication frame transmitted between the Diagnostica Stago STA Compact Max and AILS LIS middleware:

<STX>1H|\^&|||Stago STA Compact Max|||||||P|1394-97|20260815165000<CR><ETX>4F<CR><LF>
<STX>2P|1||||Price^Thomas|||M<CR><ETX>2E<CR><LF>
<STX>3O|1|ACC2026081560||^^^HEMOSTASIS|||||||A<CR><ETX>5B<CR><LF>
<STX>4R|1|^^^TP|12.1|s|11.0-13.5|N||F<CR><ETX>18<CR><LF>
<STX>5R|2|^^^INR|1.00|Ratio|0.8-1.2|N||F<CR><ETX>27<CR><LF>
<STX>6R|3|^^^TCA|31.0|s|25.0-35.0|N||F<CR><ETX>23<CR><LF>
<STX>7R|4|^^^FIB|280.0|mg/dL|200.0-400.0|N||F<CR><ETX>3A<CR><LF>
<STX>8L|1|N<CR><ETX>0A<CR><LF>

4. Diagnostica Stago STA Compact Max Clinical Parameter LOINC & Reference Range Dictionary

Instrument Code LOINC Code Clinical Parameter Name Standard Units Default Reference Range
TP (PT) 5902-2 Prothrombin Time (Temps de Prothrombine) s 11.0 - 13.5
INR 6301-6 INR (Ratio Normalisé International) Ratio 0.8 - 1.2
TCA (APTT) 3173-2 APTT (Temps de Céphaline Activée) s 25.0 - 35.0
FIB 3255-7 Fibrinogen Concentration mg/dL 200.0 - 400.0

5. Diagnostica Stago Biomedical Engineering Field Notes & Gotchas

1

In Stago operating software, configure Host Transmission to ASTM E1394 standard format.

2

Mechanical clot detection is unaffected by lipemic, icteric, or hemolyzed samples.

3

Check French test code mappings (e.g. TP for PT, TCA for APTT) in the AILS dictionary.

TECHNICAL FAQ

Frequently Asked Questions: Diagnostica Stago STA Compact Max LIS Interfacing

Common biomedical engineering questions regarding connection, worklist setup, and data capture with AILS Cloud LIMS.

Q: What communication protocol is used by Diagnostica Stago STA Compact?

Diagnostica Stago STA Compact and STA R Max use ASTM E1381/E1394 standard protocol over RS-232 serial or network data link.

Q: Why is Stago mechanical clot detection unique for LIS integration?

Stago's viscosity-based mechanical clot detection uses an oscillating magnetic steel ball, ensuring reliable clotting times even in severely lipemic or hemolyzed plasma without optical interference flags.

Q: How does AILS handle French test abbreviations from Stago analyzers?

AILS includes built-in cross-reference aliases mapping Stago French parameter codes (TP, TCA, TT) to international LOINC standards (PT, APTT, Thrombin Time).

Q: Does Stago STA Compact support bi-directional barcode worklists?

Yes. In bi-directional mode, Stago queries sample tube barcodes and downloads assigned hemostasis profiles.

Q: What serial cable is required for Stago STA Compact?

A standard DB9 Female Null Modem serial cable (Pins 2-3, 3-2, 5-5) at 9600 baud, 8 data bits, Parity None, 1 stop bit.

INSTRUMENT DIRECTORY

Related Coagulation Instrument Guides

Compare protocols, pinouts, and parameter mapping dictionaries for comparable diagnostic equipment.