Radiometer ABL800 Flex / ABL90 FLEX PLUS Bi-Directional LIS Interfacing Guide

Complete engineering specification, HL7/ASTM protocols, pinouts, and parameter mapping for Radiometer ABL800 Flex / ABL90 FLEX PLUS on AILS Cloud LIMS.

Radiometer ABL800 Flex / ABL90 FLEX PLUS Analyzer - AILS LIS Interfacing Guide
Manufacturer
Radiometer
Protocol
ASTM E1394 / HL7
Interface Mode
Bi-Directional
Throughput
STAT Critical Whole Blood Sampling

1. Radiometer ABL800 Flex / ABL90 FLEX PLUS Interfacing Overview & Diagnostic Workflow

The Radiometer ABL800 Flex / ABL90 FLEX PLUS is a precision diagnostic analyzer designed for clinical laboratory testing. Connecting the Radiometer ABL800 Flex / ABL90 FLEX PLUS directly to AILS Cloud LIMS eliminates manual data entry, prevents typographical identification errors, and delivers verified results directly to ordering physicians and hospital EMRs.

Bi-Directional Order Interfacing

AILS Gateway manages bi-directional ASTM E1394 / HL7 worklist handshakes with the Radiometer ABL800 Flex / ABL90 FLEX PLUS, streaming patient test requests as samples are loaded.

Quality Control & Reference Range Interlock

Results are checked against age- and gender-adjusted reference ranges, delta check baselines, and quality control (QC) Levey-Jennings rules before digital sign-off.

2. Radiometer ABL800 Flex / ABL90 FLEX PLUS RS-232 Serial Port & Pinout Configuration

Hardware & Socket Configuration

TCP/IP Ethernet (Port 5001) / RS-232 DB9 Serial Port (9600 Baud, 8-N-1). ASTM E1394 or HL7.

Protocol: ASTM E1394 / HL7
Interface Mode: Bi-Directional
Default Speed: 9600 / TCP 5001
// RS-232 DB9 NULL-MODEM PINOUT SERIAL COM
[Instrument DB9]                   [AILS Gateway PC]
Pin 2 (TxD / Data Out) ---------> Pin 2 (RxD / Data In)
Pin 3 (RxD / Data In)  <--------- Pin 3 (TxD / Data Out)
Pin 5 (GND / Signal)   <========> Pin 5 (GND / Signal)
Pin 7 (RTS) & Pin 8 (CTS)        [Loopback or None]
Baud: 9600 / TCP 5001 | Data Bits: 8 | Parity: None | Stop: 1
Verified with AILS Gateway Daemon & Real-Time Socket Router.

3. Radiometer ABL800 Flex / ABL90 FLEX PLUS HL7 v2.x Message Stream

Below is a validated communication frame transmitted between the Radiometer ABL800 Flex / ABL90 FLEX PLUS and AILS LIS middleware:

<STX>1H|\^&|||Radiometer ABL800 Flex|||||||P|1394-97|20260815173000<CR><ETX>5C<CR><LF>
<STX>2P|1||||Miller^George|||M<CR><ETX>2A<CR><LF>
<STX>3O|1|ACC2026081576||^^^ABG_COOX|||||||A<CR><ETX>4D<CR><LF>
<STX>4R|1|^^^pH|7.380|pH|7.350-7.450|N||F<CR><ETX>18<CR><LF>
<STX>5R|2|^^^pCO2|42.0|mmHg|35.0-45.0|N||F<CR><ETX>24<CR><LF>
<STX>6R|3|^^^pO2|92.0|mmHg|80.0-100.0|N||F<CR><ETX>21<CR><LF>
<STX>7R|4|^^^HCO3|24.5|mmol/L|22.0-26.0|N||F<CR><ETX>2E<CR><LF>
<STX>8R|5|^^^sO2|97.5|%|95.0-99.0|N||F<CR><ETX>1A<CR><LF>
<STX>9R|6|^^^LAC|1.20|mmol/L|0.50-2.00|N||F<CR><ETX>28<CR><LF>
<STX>10R|7|^^^COHb|1.1|%|0.0-2.0|N||F<CR><ETX>19<CR><LF>
<STX>11L|1|N<CR><ETX>0A<CR><LF>

4. Radiometer ABL800 Flex / ABL90 FLEX PLUS Clinical Parameter LOINC & Reference Range Dictionary

Instrument Code LOINC Code Clinical Parameter Name Standard Units Default Reference Range
pH 2744-1 Arterial Blood pH pH 7.350 - 7.450
pCO2 2019-8 Carbon Dioxide Partial Pressure mmHg 35.0 - 45.0
pO2 2703-7 Oxygen Partial Pressure mmHg 80.0 - 100.0
HCO3 1960-4 Bicarbonate Actual (HCO3-) mmol/L 22.0 - 26.0
sO2 2708-6 Oxygen Saturation Fractional % 95.0 - 99.0
LAC 2524-7 Lactate Whole Blood mmol/L 0.50 - 2.00
COHb 20563-3 Carboxyhemoglobin Fraction % 0.0 - 2.0

5. Radiometer Biomedical Engineering Field Notes & Gotchas

1

In ABL800 communication settings, configure LIS Protocol to ASTM E1394 or HL7 Standard on Port 5001.

2

Check patient temperature correction mode (pH(T), pCO2(T), pO2(T)) if entering body temperature.

3

Critical blood gas panic values (e.g. pH < 7.20 or Lactate > 4.0 mmol/L) trigger immediate ICU emergency notifications.

TECHNICAL FAQ

Frequently Asked Questions: Radiometer ABL800 Flex / ABL90 FLEX PLUS LIS Interfacing

Common biomedical engineering questions regarding connection, worklist setup, and data capture with AILS Cloud LIMS.

Q: What protocols are supported by Radiometer ABL800 Flex and ABL90 FLEX PLUS?

Radiometer blood gas analyzers support ASTM E1381/E1394 and HL7 over TCP/IP Ethernet (Port 5001) or RS-232 serial connection.

Q: How does AILS handle calculated blood gas parameters (Base Excess, Standard Bicarbonate, Alveolar-Arterial Gradient)?

AILS automatically verifies and stores all transmitted derived parameters including Base Excess (cBase), Anion Gap, and AaDO2 alongside measured primary analytes.

Q: How does optical co-oximetry synchronization work on ABL800?

Total Hemoglobin (ctHb), Oxyhemoglobin (sO2), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), and Deoxyhemoglobin (HHb) measured via 128-wavelength spectrophotometry are parsed into dedicated LOINC fields.

Q: Can Radiometer analyzers receive patient demographics via barcode scan?

Yes. When a blood gas syringe barcode is scanned, AILS Gateway sends patient ID, name, ward location, and FiO2 ventilator settings to the analyzer.

Q: What serial cable is required for ABL800 serial interfacing?

A standard DB9 Female Null Modem serial cable (Pins 2-3, 3-2, 5-5) configured at 9600 baud, 8 data bits, Parity None, 1 stop bit.

INSTRUMENT DIRECTORY

Related Blood Gas (ABG) Instrument Guides

Compare protocols, pinouts, and parameter mapping dictionaries for comparable diagnostic equipment.

View All 50+ Analyzers
PRE-CONFIGURED DRIVER READY

Connect Your Radiometer ABL800 Flex / ABL90 FLEX PLUS to AILS in Under 15 Minutes

Our specialized biomedical engineering team provides complete plug-and-play drivers, cable diagrams, and bi-directional test verification for the Radiometer ABL800 Flex / ABL90 FLEX PLUS.