URIT MEDICAL ANALYZER INTERFACING GUIDE

URIT 5250 Bi-Directional LIS Interfacing Guide

Complete engineering specification, HL7/ASTM protocols, pinouts, and parameter mapping for URIT 5250 on AILS Cloud LIMS.

URIT 5250 Analyzer - AILS LIS Interfacing Guide
Manufacturer
URIT Medical
Protocol
HL7 v2.3.1 / TCP-IP
Interface Mode
Uni-Directional
Throughput
60 samples/hour

1. URIT 5250 Interfacing Overview & Diagnostic Workflow

The URIT 5250 is an automated hematology platform engineered for high-precision complete blood counts (CBC) and white blood cell differential analysis. In modern clinical laboratory workflows, interfacing the URIT 5250 directly to AILS Cloud LIMS eliminates manual transcription delays, captures full WBC scattergram channel flags (such as immature granulocytes, atypical lymphocytes, and platelet clumps), and validates patient results against historical delta check baselines in real time.

Real-Time EDTA Tube Accessioning

AILS Gateway parses sample accession barcodes from the URIT 5250 data stream, linking multi-parameter CBC results directly to open patient encounter records without manual intervention.

Cytometry Flagging & Delta Check Interlock

Raw result segments are evaluated instantly against laboratory reference intervals and Westgard multirule delta check algorithms. Critical alerts (e.g. severe thrombocytopenia or extreme leukocytosis) trigger automatic hold flags for pathologist microscopic smear review.

2. URIT 5250 TCP/IP Socket & Network Configuration

Hardware & Socket Configuration

Ethernet RJ45 TCP/IP MLLP socket (Port 5001). Static IP configuration.

Protocol: HL7 v2.3.1 / TCP-IP
Interface Mode: Uni-Directional
Default Speed: N/A (Ethernet)
// TCP/IP MLLP SOCKET WRAPPER NETWORK
[Instrument IP Client]            [AILS Gateway Server]
Socket Address   ------------->  Static IP : Port 5001
Start of Block   ------------->  <VT> ASCII 0x0B
End of Block     ------------->  <FS><CR> ASCII 0x1C 0x0D
Payload Standard ------------->  HL7 v2.3.1 / TCP-IP
KeepAlive Probe  ------------->  Bi-Directional ACK (AA)
Verified with AILS Gateway Daemon & Real-Time Socket Router.

3. URIT 5250 HL7 v2.x Message Stream

Below is a validated communication frame transmitted between the URIT 5250 and AILS LIS middleware:

MSH|^~\&|URIT5250|URIT|AILS_LIMS|LAB_MAIN|20260815153000||ORU^R01|MSG009960|P|2.3.1
PID|1||P982420||Cooper^James||19840718|M
OBR|1|ORD982420|ACC2026081538|CBC5D^5-Part CBC|||20260815152500
OBX|1|NM|WBC^White Blood Cells||7.45|10*3/uL|4.0-10.0|N|||F
OBX|2|NM|HGB^Hemoglobin||14.8|g/dL|13.0-17.0|N|||F
OBX|3|NM|PLT^Platelets||260|10*3/uL|150-450|N|||F
OBX|4|NM|NEU%^Neutrophils||62.8|%|40.0-75.0|N|||F

4. URIT 5250 Clinical Parameter LOINC & Reference Range Dictionary

Instrument Code LOINC Code Clinical Parameter Name Standard Units Default Reference Range
WBC 6690-2 White Blood Cells 10*3/uL 4.0 - 10.0
HGB 718-7 Hemoglobin g/dL 13.0 - 17.0
PLT 777-3 Platelets 10*3/uL 150 - 450

5. URIT Medical Biomedical Engineering Field Notes & Gotchas

1

In URIT 5250 system settings, configure Communication Protocol to HL7 TCP.

2

AILS Gateway returns ACK^R01 confirmation to complete the transmission handshake.

3

Verify that laser optical scattergram data is enabled in data transmission profile.

TECHNICAL FAQ

Frequently Asked Questions: URIT 5250 LIS Interfacing

Common biomedical engineering questions regarding connection, worklist setup, and data capture with AILS Cloud LIMS.

Q: What communication protocol does URIT 5250 use?

URIT 5250 uses HL7 v2.3.1 over TCP/IP MLLP socket.

Q: How does AILS Gateway process 5-part differential results from URIT 5250?

Full 5-part differential parameters (NEU, LYM, MONO, EOS, BASO) are extracted from OBX segments and mapped to LOINC codes.

Q: What network configuration is required for URIT 5250?

Connect via Cat6 Ethernet cable and set AILS Gateway to listen on TCP Port 5001.

Q: Does URIT 5250 support bi-directional worklists?

Yes. Sample tube barcodes are queried against AILS LIMS and test profiles are downloaded automatically.

Q: How are abnormal cell warnings handled on URIT 5250?

Flags such as Immature Granulocytes and Blast alerts trigger automatic pathologist review holds.

INSTRUMENT DIRECTORY

Related Hematology Instrument Guides

Compare protocols, pinouts, and parameter mapping dictionaries for comparable diagnostic equipment.

View All 50+ Analyzers
PRE-CONFIGURED DRIVER READY

Connect Your URIT 5250 to AILS in Under 15 Minutes

Our specialized biomedical engineering team provides complete plug-and-play drivers, cable diagrams, and bi-directional test verification for the URIT 5250.