URIT MEDICAL ANALYZER INTERFACING GUIDE

URIT 5250 Bi-Directional LIS Interfacing Guide

Complete engineering specification, HL7/ASTM protocols, pinouts, and parameter mapping for URIT 5250 on AILS Cloud LIMS.

URIT 5250 Analyzer - AILS LIS Interfacing Guide
Manufacturer
URIT Medical
Protocol
HL7 v2.3.1 / TCP-IP
Interface Mode
Uni-Directional
Throughput
60 samples/hour

1. Instrument Overview & Interfacing Architecture

The URIT 5250 is a high-performance hematology instrument designed for diagnostic laboratories and hospital health networks. Integrating this analyzer directly with AILS Cloud LIMS eliminates manual transcription delays, prevents typographical patient identification errors, and automatically enforces Westgard delta check algorithms.

Bi-Directional Order Push

AILS automatically generates worklist queries (QRY^Q02 / ASTM orders) mapped to patient accession barcodes and pushes test instructions directly to the instrument sample tray.

Real-Time Result Ingestion

Raw OBX / Result records are parsed instantly within 200 milliseconds, flagged for abnormal reference ranges, and linked to pathologist digital signature queues.

2. Hardware Connection & Communication Settings

// Physical & Socket Configuration
Ethernet RJ45 TCP/IP MLLP socket (Port 5001). Static IP configuration.
// Protocol: HL7 v2.3.1 / TCP-IP | Connection Type: TCP/IP MLLP (Port 5001)

3. Sample Message Wire Frame

Below is a real-world communication frame transmitted between the URIT 5250 and AILS LIS middleware:

MSH|^~\&|URIT5250|URIT|AILS_LIMS|LAB_MAIN|20260815153000||ORU^R01|MSG009960|P|2.3.1
PID|1||P982420||Cooper^James||19840718|M
OBR|1|ORD982420|ACC2026081538|CBC5D^5-Part CBC|||20260815152500
OBX|1|NM|WBC^White Blood Cells||7.45|10*3/uL|4.0-10.0|N|||F
OBX|2|NM|HGB^Hemoglobin||14.8|g/dL|13.0-17.0|N|||F
OBX|3|NM|PLT^Platelets||260|10*3/uL|150-450|N|||F
OBX|4|NM|NEU%^Neutrophils||62.8|%|40.0-75.0|N|||F

4. Parameter & LOINC Standard Code Mapping

Instrument Code LOINC Code Clinical Parameter Name Standard Units Default Reference Range
WBC 6690-2 White Blood Cells 10*3/uL 4.0 - 10.0
HGB 718-7 Hemoglobin g/dL 13.0 - 17.0
PLT 777-3 Platelets 10*3/uL 150 - 450

5. Biomedical Engineering Tips & Common Gotchas

1

In URIT 5250 system settings, configure Communication Protocol to HL7 TCP.

2

AILS Gateway returns ACK^R01 confirmation to complete the transmission handshake.

3

Verify that laser optical scattergram data is enabled in data transmission profile.

TECHNICAL FAQ

Frequently Asked Questions: URIT 5250 Interfacing

Common biomedical engineering questions regarding connection, worklist setup, and data capture with AILS Cloud LIMS.

Q: What communication protocol does URIT 5250 use?

URIT 5250 uses HL7 v2.3.1 over TCP/IP MLLP socket.

Q: How does AILS Gateway process 5-part differential results from URIT 5250?

Full 5-part differential parameters (NEU, LYM, MONO, EOS, BASO) are extracted from OBX segments and mapped to LOINC codes.

Q: What network configuration is required for URIT 5250?

Connect via Cat6 Ethernet cable and set AILS Gateway to listen on TCP Port 5001.

Q: Does URIT 5250 support bi-directional worklists?

Yes. Sample tube barcodes are queried against AILS LIMS and test profiles are downloaded automatically.

Q: How are abnormal cell warnings handled on URIT 5250?

Flags such as Immature Granulocytes and Blast alerts trigger automatic pathologist review holds.