Sysmex XT-2000i Bi-Directional LIS Interfacing Guide

Complete engineering specification, HL7/ASTM protocols, pinouts, and parameter mapping for Sysmex XT-2000i on AILS Cloud LIMS.

Sysmex XT-2000i Analyzer - AILS LIS Interfacing Guide
Manufacturer
Sysmex
Protocol
ASTM E1394 / E1381
Interface Mode
Uni-Directional / Bi-Directional
Throughput
80 samples/hour

1. Sysmex XT-2000i Interfacing Overview & Diagnostic Workflow

The Sysmex XT-2000i is an automated hematology platform engineered for high-precision complete blood counts (CBC) and white blood cell differential analysis. In modern clinical laboratory workflows, interfacing the Sysmex XT-2000i directly to AILS Cloud LIMS eliminates manual transcription delays, captures full WBC scattergram channel flags (such as immature granulocytes, atypical lymphocytes, and platelet clumps), and validates patient results against historical delta check baselines in real time.

Automated Barcode Sample Worklist

AILS Gateway coordinates with the Sysmex XT-2000i auto-sampler and barcode scanner, dispatching ASTM order records mapped to patient accession IDs to queue 3-part or 5-part differential testing.

Cytometry Flagging & Delta Check Interlock

Raw result segments are evaluated instantly against laboratory reference intervals and Westgard multirule delta check algorithms. Critical alerts (e.g. severe thrombocytopenia or extreme leukocytosis) trigger automatic hold flags for pathologist microscopic smear review.

2. Sysmex XT-2000i RS-232 Serial Port & Pinout Configuration

Hardware & Socket Configuration

TCP/IP Ethernet (Port 5001) or RS-232 Serial (9600 Baud, 8-N-1).

Protocol: ASTM E1394 / E1381
Interface Mode: Uni-Directional / Bi-Directional
Default Speed: 9600 / 19200
// RS-232 DB9 NULL-MODEM PINOUT SERIAL COM
[Instrument DB9]                   [AILS Gateway PC]
Pin 2 (TxD / Data Out) ---------> Pin 2 (RxD / Data In)
Pin 3 (RxD / Data In)  <--------- Pin 3 (TxD / Data Out)
Pin 5 (GND / Signal)   <========> Pin 5 (GND / Signal)
Pin 7 (RTS) & Pin 8 (CTS)        [Loopback or None]
Baud: 9600 / 19200 | Data Bits: 8 | Parity: None | Stop: 1
Verified with AILS Gateway Daemon & Real-Time Socket Router.

3. Sysmex XT-2000i ASTM E1394 Record Frame

Below is a validated communication frame transmitted between the Sysmex XT-2000i and AILS LIS middleware:

<STX>1H|\^&|||Sysmex XT-2000i|||||||P|1394-97|20260815124000<CR><ETX>5C<CR><LF>
<STX>2P|1||||Williams^David|||M<CR><ETX>3E<CR><LF>
<STX>3O|1|ACC2026081504||^^^CBC+RET|||||||A<CR><ETX>4D<CR><LF>
<STX>4R|1|^^^WBC|7.40|10*3/uL|4.0-11.0|N||F<CR><ETX>1B<CR><LF>
<STX>5R|2|^^^RET%|1.25|%|0.5-2.0|N||F<CR><ETX>2C<CR><LF>
<STX>6R|3|^^^IRF|0.22|Ratio|0.1-0.3|N||F<CR><ETX>3A<CR><LF>
<STX>7L|1|N<CR><ETX>0A<CR><LF>

4. Sysmex XT-2000i Clinical Parameter LOINC & Reference Range Dictionary

Instrument Code LOINC Code Clinical Parameter Name Standard Units Default Reference Range
WBC 6690-2 White Blood Count 10*3/uL 4.0 - 11.0
RET% 17849-1 Reticulocyte Percentage % 0.5 - 2.0
RET# 787-2 Reticulocyte Absolute 10*6/uL 0.02 - 0.10
IRF 33256-9 Immature Reticulocyte Fraction Ratio 0.1 - 0.3

5. Sysmex Biomedical Engineering Field Notes & Gotchas

1

Ensure the fluorescent optical bench reticulocyte channel is enabled in analyzer transmission settings.

2

Verify that sample rack barcode reader reads both Codabar and Code 128 formats.

3

AILS Gateway decodes extended platelet indices (P-LCR, MPV, PDW) automatically.

TECHNICAL FAQ

Frequently Asked Questions: Sysmex XT-2000i LIS Interfacing

Common biomedical engineering questions regarding connection, worklist setup, and data capture with AILS Cloud LIMS.

Q: What protocol does Sysmex XT-2000i use for data transmission?

Sysmex XT-2000i uses ASTM E1394 standard protocol over TCP/IP socket (Port 5001) or RS-232 serial communication.

Q: How are optical reticulocyte parameters mapped in AILS?

Parameters including RET#, RET%, IRF (Immature Reticulocyte Fraction), and LFR/MFR/HFR are parsed into dedicated LOINC coded fields.

Q: Can XT-2000i connect in bi-directional mode?

Yes. When connected via TCP/IP to AILS Gateway, the analyzer queries the host upon reading sample barcodes and receives test panel instructions.

Q: How does AILS handle flagging on Sysmex XT-2000i?

Abnormal WBC scattergram flags, RBC agglutination warnings, and platelet clump flags are recorded in the accession audit log.

Q: What is the recommended network setup for Sysmex XT-2000i?

Assign a static IP address to the Sysmex IPU PC and set AILS Gateway to listen on TCP Port 5001 as Server.

INSTRUMENT DIRECTORY

Related Hematology Instrument Guides

Compare protocols, pinouts, and parameter mapping dictionaries for comparable diagnostic equipment.

View All 50+ Analyzers
PRE-CONFIGURED DRIVER READY

Connect Your Sysmex XT-2000i to AILS in Under 15 Minutes

Our specialized biomedical engineering team provides complete plug-and-play drivers, cable diagrams, and bi-directional test verification for the Sysmex XT-2000i.