Sysmex XT-2000i Bi-Directional LIS Interfacing Guide

Complete engineering specification, HL7/ASTM protocols, pinouts, and parameter mapping for Sysmex XT-2000i on AILS Cloud LIMS.

Sysmex XT-2000i Analyzer - AILS LIS Interfacing Guide
Manufacturer
Sysmex
Protocol
ASTM E1394 / E1381
Interface Mode
Uni-Directional / Bi-Directional
Throughput
80 samples/hour

1. Instrument Overview & Interfacing Architecture

The Sysmex XT-2000i is a high-performance hematology instrument designed for diagnostic laboratories and hospital health networks. Integrating this analyzer directly with AILS Cloud LIMS eliminates manual transcription delays, prevents typographical patient identification errors, and automatically enforces Westgard delta check algorithms.

Bi-Directional Order Push

AILS automatically generates worklist queries (QRY^Q02 / ASTM orders) mapped to patient accession barcodes and pushes test instructions directly to the instrument sample tray.

Real-Time Result Ingestion

Raw OBX / Result records are parsed instantly within 200 milliseconds, flagged for abnormal reference ranges, and linked to pathologist digital signature queues.

2. Hardware Connection & Communication Settings

// Physical & Socket Configuration
TCP/IP Ethernet (Port 5001) or RS-232 Serial (9600 Baud, 8-N-1).
// Protocol: ASTM E1394 / E1381 | Connection Type: TCP/IP (Port 5001) / RS232

3. Sample Message Wire Frame

Below is a real-world communication frame transmitted between the Sysmex XT-2000i and AILS LIS middleware:

<STX>1H|\^&|||Sysmex XT-2000i|||||||P|1394-97|20260815124000<CR><ETX>5C<CR><LF>
<STX>2P|1||||Williams^David|||M<CR><ETX>3E<CR><LF>
<STX>3O|1|ACC2026081504||^^^CBC+RET|||||||A<CR><ETX>4D<CR><LF>
<STX>4R|1|^^^WBC|7.40|10*3/uL|4.0-11.0|N||F<CR><ETX>1B<CR><LF>
<STX>5R|2|^^^RET%|1.25|%|0.5-2.0|N||F<CR><ETX>2C<CR><LF>
<STX>6R|3|^^^IRF|0.22|Ratio|0.1-0.3|N||F<CR><ETX>3A<CR><LF>
<STX>7L|1|N<CR><ETX>0A<CR><LF>

4. Parameter & LOINC Standard Code Mapping

Instrument Code LOINC Code Clinical Parameter Name Standard Units Default Reference Range
WBC 6690-2 White Blood Count 10*3/uL 4.0 - 11.0
RET% 17849-1 Reticulocyte Percentage % 0.5 - 2.0
RET# 787-2 Reticulocyte Absolute 10*6/uL 0.02 - 0.10
IRF 33256-9 Immature Reticulocyte Fraction Ratio 0.1 - 0.3

5. Biomedical Engineering Tips & Common Gotchas

1

Ensure the fluorescent optical bench reticulocyte channel is enabled in analyzer transmission settings.

2

Verify that sample rack barcode reader reads both Codabar and Code 128 formats.

3

AILS Gateway decodes extended platelet indices (P-LCR, MPV, PDW) automatically.

TECHNICAL FAQ

Frequently Asked Questions: Sysmex XT-2000i Interfacing

Common biomedical engineering questions regarding connection, worklist setup, and data capture with AILS Cloud LIMS.

Q: What protocol does Sysmex XT-2000i use for data transmission?

Sysmex XT-2000i uses ASTM E1394 standard protocol over TCP/IP socket (Port 5001) or RS-232 serial communication.

Q: How are optical reticulocyte parameters mapped in AILS?

Parameters including RET#, RET%, IRF (Immature Reticulocyte Fraction), and LFR/MFR/HFR are parsed into dedicated LOINC coded fields.

Q: Can XT-2000i connect in bi-directional mode?

Yes. When connected via TCP/IP to AILS Gateway, the analyzer queries the host upon reading sample barcodes and receives test panel instructions.

Q: How does AILS handle flagging on Sysmex XT-2000i?

Abnormal WBC scattergram flags, RBC agglutination warnings, and platelet clump flags are recorded in the accession audit log.

Q: What is the recommended network setup for Sysmex XT-2000i?

Assign a static IP address to the Sysmex IPU PC and set AILS Gateway to listen on TCP Port 5001 as Server.