Sysmex UF-4000 / UF-5000 Bi-Directional LIS Interfacing Guide

Complete engineering specification, HL7/ASTM protocols, pinouts, and parameter mapping for Sysmex UF-4000 / UF-5000 on AILS Cloud LIMS.

Sysmex UF-4000 / UF-5000 Analyzer - AILS LIS Interfacing Guide
Manufacturer
Sysmex
Protocol
ASTM E1394 / HL7 v2.5
Interface Mode
Bi-Directional
Throughput
105 samples/hour

1. Sysmex UF-4000 / UF-5000 Interfacing Overview & Diagnostic Workflow

The Sysmex UF-4000 / UF-5000 is a precision diagnostic analyzer designed for clinical laboratory testing. Connecting the Sysmex UF-4000 / UF-5000 directly to AILS Cloud LIMS eliminates manual data entry, prevents typographical identification errors, and delivers verified results directly to ordering physicians and hospital EMRs.

Bi-Directional Order Interfacing

AILS Gateway manages bi-directional ASTM E1394 / HL7 v2.5 worklist handshakes with the Sysmex UF-4000 / UF-5000, streaming patient test requests as samples are loaded.

Quality Control & Reference Range Interlock

Results are checked against age- and gender-adjusted reference ranges, delta check baselines, and quality control (QC) Levey-Jennings rules before digital sign-off.

2. Sysmex UF-4000 / UF-5000 TCP/IP Socket & Network Configuration

Hardware & Socket Configuration

Ethernet RJ45 TCP/IP MLLP Socket (Port 5001). ASTM E1394 or HL7 v2.5.

Protocol: ASTM E1394 / HL7 v2.5
Interface Mode: Bi-Directional
Default Speed: N/A (TCP 5001)
// TCP/IP MLLP SOCKET WRAPPER NETWORK
[Instrument IP Client]            [AILS Gateway Server]
Socket Address   ------------->  Static IP : Port 5001
Start of Block   ------------->  <VT> ASCII 0x0B
End of Block     ------------->  <FS><CR> ASCII 0x1C 0x0D
Payload Standard ------------->  ASTM E1394 / HL7 v2.5
KeepAlive Probe  ------------->  Bi-Directional ACK (AA)
Verified with AILS Gateway Daemon & Real-Time Socket Router.

3. Sysmex UF-4000 / UF-5000 HL7 v2.x Message Stream

Below is a validated communication frame transmitted between the Sysmex UF-4000 / UF-5000 and AILS LIS middleware:

<STX>1H|\^&|||Sysmex UF-5000|||||||P|1394-97|20260815170000<CR><ETX>4B<CR><LF>
<STX>2P|1||||Sanchez^Elena|||F<CR><ETX>2A<CR><LF>
<STX>3O|1|ACC2026081562||^^^URINE_SEDIMENT|||||||A<CR><ETX>5F<CR><LF>
<STX>4R|1|^^^RBC|12.4|/uL|0.0-15.0|N||F<CR><ETX>1D<CR><LF>
<STX>5R|2|^^^WBC|45.2|/uL|0.0-25.0|H||F<CR><ETX>2C<CR><LF>
<STX>6R|3|^^^EC|8.0|/uL|0.0-10.0|N||F<CR><ETX>17<CR><LF>
<STX>7R|4|^^^CAST|0.45|/uL|0.0-1.0|N||F<CR><ETX>2E<CR><LF>
<STX>8R|5|^^^BACT|850.0|/uL|0.0-100.0|H||F<CR><ETX>34<CR><LF>
<STX>9L|1|N<CR><ETX>0A<CR><LF>

4. Sysmex UF-4000 / UF-5000 Clinical Parameter LOINC & Reference Range Dictionary

Instrument Code LOINC Code Clinical Parameter Name Standard Units Default Reference Range
RBC 20448-7 Red Blood Cells Quantitative /uL 0.0 - 15.0
WBC 20449-5 White Blood Cells Quantitative /uL 0.0 - 25.0
EC 20450-3 Epithelial Cells Quantitative /uL 0.0 - 10.0
CAST 20451-1 Hyaline & Pathological Casts /uL 0.0 - 1.0
BACT 20452-9 Bacteria Count (Bacteriuria) /uL 0.0 - 100.0

5. Sysmex Biomedical Engineering Field Notes & Gotchas

1

In UF-5000 IPU communication settings, configure Host Interface to ASTM E1394 over TCP Port 5001.

2

High bacterial counts (BACT > 100/uL) with elevated WBCs automatically flag suspected Urinary Tract Infection (UTI).

3

Dysmorphic RBC information (Non-Lysed RBC, Glomerular vs Non-Glomerular) is captured in dedicated clinical research fields.

TECHNICAL FAQ

Frequently Asked Questions: Sysmex UF-4000 / UF-5000 LIS Interfacing

Common biomedical engineering questions regarding connection, worklist setup, and data capture with AILS Cloud LIMS.

Q: What technology is used by Sysmex UF-5000 for urinalysis?

Sysmex UF-5000 uses blue semiconductor laser fluorescence flow cytometry to count and classify urinary formed elements (RBC, WBC, casts, epithelial cells, bacteria, yeast, and crystals) in quantitative units per microliter (/uL).

Q: How does UF-5000 transmit results to AILS LIMS?

UF-5000 transmits complete particle enumeration data and scattergram flags over TCP/IP socket (Port 5001) using ASTM E1394 or HL7 v2.5.

Q: How does AILS handle automated bacteriuria and UTI screening?

When BACT and WBC counts exceed clinical thresholds, AILS automatically flags the specimen for reflex urine culture and antibiotic susceptibility testing (AST).

Q: Can Sysmex UF-5000 connect alongside a urine test strip analyzer (e.g. UC-3500)?

Yes. The consolidated UN-Series urine workcell combines test strip chemistry and flow cytometry sediment analysis into a single unified patient report.

Q: How are abnormal casts and dysmorphic RBCs reviewed in AILS?

Pathological casts and abnormal red cell morphology flags automatically route the case to the pathologist microscopy review queue.

INSTRUMENT DIRECTORY

Related Urinalysis Instrument Guides

Compare protocols, pinouts, and parameter mapping dictionaries for comparable diagnostic equipment.

View All 50+ Analyzers
PRE-CONFIGURED DRIVER READY

Connect Your Sysmex UF-4000 / UF-5000 to AILS in Under 15 Minutes

Our specialized biomedical engineering team provides complete plug-and-play drivers, cable diagrams, and bi-directional test verification for the Sysmex UF-4000 / UF-5000.