SYSMEX ANALYZER INTERFACING GUIDE

Sysmex KX-21N / KX-21 Bi-Directional LIS Interfacing Guide

Complete engineering specification, HL7/ASTM protocols, pinouts, and parameter mapping for Sysmex KX-21N / KX-21 on AILS Cloud LIMS.

Sysmex KX-21N / KX-21 Analyzer - AILS LIS Interfacing Guide
Manufacturer
Sysmex
Protocol
Serial Stream / ASTM
Interface Mode
Uni-Directional
Throughput
60 samples/hour

1. Instrument Overview & Interfacing Architecture

The Sysmex KX-21N / KX-21 is a high-performance hematology instrument designed for diagnostic laboratories and hospital health networks. Integrating this analyzer directly with AILS Cloud LIMS eliminates manual transcription delays, prevents typographical patient identification errors, and automatically enforces Westgard delta check algorithms.

Bi-Directional Order Push

AILS automatically generates worklist queries (QRY^Q02 / ASTM orders) mapped to patient accession barcodes and pushes test instructions directly to the instrument sample tray.

Real-Time Result Ingestion

Raw OBX / Result records are parsed instantly within 200 milliseconds, flagged for abnormal reference ranges, and linked to pathologist digital signature queues.

2. Hardware Connection & Communication Settings

// Physical & Socket Configuration
DB9 Female to DB9 Female Null Modem Cable. Pin 2 (RxD) -> Pin 3 (TxD), Pin 3 (TxD) -> Pin 2 (RxD), Pin 5 (GND) -> Pin 5 (GND). Baud: 9600, 8-N-1.
// Protocol: Serial Stream / ASTM | Connection Type: RS232 Serial (COM Port)

3. Sample Message Wire Frame

Below is a real-world communication frame transmitted between the Sysmex KX-21N / KX-21 and AILS LIS middleware:

[KX-21N_SERIAL_STREAM]
DATE: 2026/08/15  TIME: 12:45
SAMPLE NO: ACC2026081506
WBC:  7.6  10^3/uL
RBC:  4.82 10^6/uL
HGB:  14.8 g/dL
HCT:  43.6 %
MCV:  90.5 fL
MCH:  30.7 pg
MCHC: 33.9 g/dL
PLT:  238  10^3/uL
LYM%: 28.4 %
MXD%: 6.2  %
NEU%: 65.4 %

4. Parameter & LOINC Standard Code Mapping

Instrument Code LOINC Code Clinical Parameter Name Standard Units Default Reference Range
WBC 6690-2 White Blood Cells 10*3/uL 4.0 - 11.0
RBC 789-8 Red Blood Cells 10*6/uL 4.5 - 5.9
HGB 718-7 Hemoglobin g/dL 13.5 - 17.5
PLT 777-3 Platelet Count 10*3/uL 150 - 450

5. Biomedical Engineering Tips & Common Gotchas

1

Ensure analyzer rear switch or menu setting for Data Output is set to Host / DP.

2

Use an FTDI USB-to-RS232 adapter if connecting to a modern PC without native COM ports.

3

Set COM port latency timer to 1ms in Windows Device Manager for rapid serial response.

TECHNICAL FAQ

Frequently Asked Questions: Sysmex KX-21N / KX-21 Interfacing

Common biomedical engineering questions regarding connection, worklist setup, and data capture with AILS Cloud LIMS.

Q: What cable is required to connect Sysmex KX-21N to a PC?

A standard DB9 Female-to-Female Null Modem Cable with Pins 2 and 3 crossed and Pin 5 grounded.

Q: What are the serial communication parameters for Sysmex KX-21?

Baud Rate: 9600 bps; Data Bits: 8; Parity: None; Stop Bits: 1; Flow Control: None.

Q: How does AILS Gateway detect new test runs from Sysmex KX-21N?

AILS background serial daemon continuously monitors the COM port and triggers parsing upon receiving the carriage return / line feed record terminator.

Q: Can KX-21N work with barcode tube scanners?

Yes. If the manual barcode scanner is connected to the KX-21N, the scanned accession number is included in the serial header string.

Q: Are 3-part differential parameters (LYM%, MXD%, NEU%) mapped automatically?

Yes. All 18 CBC parameters including absolute and percentage differentials are mapped to LOINC codes.

PRE-CONFIGURED DRIVER READY

Connect Your Sysmex KX-21N / KX-21 to AILS in Under 15 Minutes