VITAL SCIENTIFIC ANALYZER INTERFACING GUIDE

Selectra ProS Bi-Directional LIS Interfacing Guide

Complete engineering specification, HL7/ASTM protocols, pinouts, and parameter mapping for Selectra ProS on AILS Cloud LIMS.

Selectra ProS Analyzer - AILS LIS Interfacing Guide
Manufacturer
Vital Scientific
Protocol
ASTM E1381 / E1394
Interface Mode
Uni-Directional / Bi-Directional
Throughput
140 tests/hour

1. Instrument Overview & Interfacing Architecture

The Selectra ProS is a high-performance biochemistry instrument designed for diagnostic laboratories and hospital health networks. Integrating this analyzer directly with AILS Cloud LIMS eliminates manual transcription delays, prevents typographical patient identification errors, and automatically enforces Westgard delta check algorithms.

Bi-Directional Order Push

AILS automatically generates worklist queries (QRY^Q02 / ASTM orders) mapped to patient accession barcodes and pushes test instructions directly to the instrument sample tray.

Real-Time Result Ingestion

Raw OBX / Result records are parsed instantly within 200 milliseconds, flagged for abnormal reference ranges, and linked to pathologist digital signature queues.

2. Hardware Connection & Communication Settings

// Physical & Socket Configuration
TCP/IP Ethernet (Port 5002) / RS-232 DB9 Serial Port (9600 Baud, 8-N-1).
// Protocol: ASTM E1381 / E1394 | Connection Type: TCP/IP (Port 5002) & RS232

3. Sample Message Wire Frame

Below is a real-world communication frame transmitted between the Selectra ProS and AILS LIS middleware:

<STX>1H|\^&|||Selectra ProS|||||||P|1394-97|20260815141000<CR><ETX>4B<CR><LF>
<STX>2P|1||||Clark^Susan|||F<CR><ETX>2A<CR><LF>
<STX>3O|1|ACC2026081518||^^^BIOCHEM|||||||A<CR><ETX>4C<CR><LF>
<STX>4R|1|^^^GLU|98.5|mg/dL|70.0-100.0|N||F<CR><ETX>1A<CR><LF>
<STX>5R|2|^^^CHOL|188.0|mg/dL|140.0-200.0|N||F<CR><ETX>2B<CR><LF>
<STX>6L|1|N<CR><ETX>0A<CR><LF>

4. Parameter & LOINC Standard Code Mapping

Instrument Code LOINC Code Clinical Parameter Name Standard Units Default Reference Range
GLU 2345-7 Glucose Fasting mg/dL 70.0 - 100.0
CHOL 2093-3 Total Cholesterol mg/dL 140.0 - 200.0

5. Biomedical Engineering Tips & Common Gotchas

1

Configure Selectra ProS to communicate via TCP port 5002 in Host Settings.

2

Check that reagent barcoded bottles are scanned before launching analytical runs.

3

Ensure AILS Gateway is configured with matching STX/ETX framing.

TECHNICAL FAQ

Frequently Asked Questions: Selectra ProS Interfacing

Common biomedical engineering questions regarding connection, worklist setup, and data capture with AILS Cloud LIMS.

Q: What interface protocols are supported by Selectra ProS?

Selectra ProS supports ASTM E1381/E1394 over TCP/IP socket (Port 5002) and RS-232 serial.

Q: How does Selectra ProS transmit biochemistry test batches?

Results are transmitted in real time as ASTM record frames as each photometric reading completes.

Q: Can Selectra ProS operate with bi-directional worklists?

Yes. Sample barcodes scanned on the rotor are queried against AILS LIMS to fetch test profiles.

Q: How are ISE module parameters mapped on Selectra ProS?

Sodium, Potassium, and Chloride values are mapped to dedicated LOINC codes in AILS LIMS.

Q: What baud rate is used for Selectra ProS serial connections?

Standard baud rate is 9600 bps, 8 data bits, no parity, 1 stop bit.

PRE-CONFIGURED DRIVER READY

Connect Your Selectra ProS to AILS in Under 15 Minutes

Our specialized biomedical engineering team provides complete plug-and-play drivers, cable diagrams, and bi-directional test verification for the Selectra ProS.