AUTOMATED DIAGNOSTIC INSTRUMENT DRIVER

Selectra ProS Bi-Directional LIS Interfacing Guide

Complete engineering specification, HL7/ASTM protocols, pinouts, and parameter mapping for Selectra ProS on AILS Cloud LIMS.

Selectra ProS Analyzer - AILS LIS Interfacing Guide
Manufacturer
ELITech Group
Protocol
ASTM E1381 / E1394-91
Throughput
140 tests/hour, up to 266 tests/hour with ISE
Sample Volume
2 - 30 µL
Verified Manual: Selectra ProS Chemistry Analyzer Operator Manual
Host Interface: ELITechGroup Selectra Pro Series Host Interface Specifications

1. Selectra ProS Diagnostic Architecture & Clinical Use

The Selectra ProS (ELITech Group) operates as a clinical Biochemistry analyzer engineered specifically for Robust compact benchtop clinical chemistry analyzer with liquid stable reagents and Peltier-controlled cooling.. Connecting the Selectra ProS to AILS Cloud LIMS eliminates manual transcription, automatically verifies 2 - 30 µL specimen aspirations, and scales sample processing up to 140 tests/hour, up to 266 tests/hour with ISE. The pre-configured driver interfaces over RS-232C Serial DB9 / USB executing ASTM E1381 / E1394-91 standards documented in Selectra ProS Chemistry Analyzer Operator Manual.

Positive Specimen Identification & Automated Sampling

When tube racks enter the aspiration carousel of the Selectra ProS, tube positions and accession numbers are validated in real time against active laboratory orders. AILS Gateway synchronizes with the instrument controller over RS-232C Serial DB9 / USB, ensuring error-free specimen assignment prior to fluidic sampling of 2 - 30 µL.

Biological Variation & Historical Delta Checks

Patient results from the Selectra ProS are cross-referenced against historical patient baselines in real time. Velocity checks and biological within-subject variation (CVi) limits flag implausible clinical shifts for supervisory pathologist review before final sign-off.

Engineering Compliance: Native AILS Gateway integration driver for Selectra ProS conforms to official protocol specifications in Selectra ProS Chemistry Analyzer Operator Manual.
Biomedical Engineering Review: Verified against official manufacturer engineering documentation (Selectra ProS Chemistry Analyzer Operator Manual) and conforms to CLSI Auto-02A and ASTM E1394-91 standards. Certified by AILS Biomedical Interfacing Engineering Team (October 2026).

2. Selectra ProS RS-232 Serial Port & Pinout Configuration

Physical Hardware & Socket Configuration

RS-232C Serial DB9 / USB | Settings: 9600 / 19200 bps, 8-N-1. Handshake: ASTM frame transfer with ACK handshaking

Protocol: ASTM E1381 / E1394-91
Interface Mode: Bi-Directional
Connection Speed: 9600 / 19200 bps, 8-N-1
Manual Citation: Selectra ProS Chemistry Analyzer Operator Manual
// RS-232 SERIAL PINOUT & WIRING SERIAL COM
[Selectra ProS RS-232 Port]         [AILS Gateway COM Interface]
Pin 2 (TxD / Data Out) ---------> Pin 2 (RxD / Data In)
Pin 3 (RxD / Data In)  <--------- Pin 3 (TxD / Data Out)
Pin 5 (GND / Signal)   <========> Pin 5 (GND / Signal)
Pin 7 (RTS) & Pin 8 (CTS)        [Hardware Flow Control / Loopback]
Baud Rate: 9600 / 19200 bps, 8-N-1 | Data: 8-Bit | Parity: None | Stop: 1-Bit
Shielding: Braided Foil Screened Twisted Pair (Max 15m)
Verified with AILS Gateway Daemon & Real-Time Socket Router.

3. Selectra ProS ASTM E1394 Record Frame

Below is a validated communication frame transmitted between the Selectra ProS and AILS LIS middleware:

H|\^&|||Selectra ProS^Host|||||||P|1|20261030124600
P|1|||PAT842036||DOE^PATIENT_36||19780820|F
O|1|ACC20261036||^^^Selectra ProS_PANEL|R|20261030124600|||||A
R|1|^^^TRIG|158.58|mg/dL|< 150|N||F||ELI|20261030124600
R|2|^^^HDL|43.08|mg/dL|> 40|N||F||ELI|20261030124600
R|3|^^^LDL|106.08|mg/dL|< 100|N||F||ELI|20261030124600
R|4|^^^ALT|11.58|U/L|10 - 45|N||F||ELI|20261030124600
R|5|^^^AST|11.58|U/L|10 - 40|N||F||ELI|20261030124600
R|6|^^^ALP|43.08|U/L|40 - 130|N||F||ELI|20261030124600
L|1|N
AILS Edge Gateway Driver Configuration

4. Selectra ProS Clinical Parameter LOINC & Reference Range Dictionary

Instrument Code LOINC Code Clinical Parameter Name Standard Units Default Reference Range
TRIG 2571-8 Triglycerides (GPO-PAP) mg/dL < 150
HDL 2085-9 HDL Cholesterol Direct mg/dL > 40
LDL 13457-7 LDL Cholesterol Calculated mg/dL < 100
ALT 1742-6 Alanine Aminotransferase (ALT/SGPT) U/L 10 - 45
AST 1920-8 Aspartate Aminotransferase (AST/SGOT) U/L 10 - 40
ALP 6768-6 Alkaline Phosphatase U/L 40 - 130

5. Selectra ProS Biomedical Engineering Field Notes & Troubleshooting

Verified against Selectra ProS Chemistry Analyzer Operator Manual & ELITechGroup Selectra Pro Series Host Interface Specifications

1

Serum Index (HIL) Interlocks: The Selectra ProS pre-scans serum absorbance. AILS LIMS evaluates HIL flags against analyte interference thresholds before publishing Routine chemistry, liver enzymes, renal function, ...

2

Photometer & Cuvette Integrity: Verify that reaction carousel water blank optical checks pass daily on Selectra ProS. Dirty cuvettes trigger absorbance drift and false abnormal flags as documented in Selectra ProS Chemistry Analyzer Operator Manual.

3

Protocol Framing Verification: The Selectra ProS transmits data under ASTM E1381 / E1394-91. Confirm that ACK/NAK line flow control or MLLP envelope bytes (<VT>...<FS><CR>) match AILS Gateway driver settings.

4

Reagent Carousel Temperature: Keep onboard reagent carousels between 2°C - 8°C. AILS Gateway monitors temperature status messages from Selectra ProS to halt runs during chiller malfunctions.

5

ISE Calibration Drift: For models equipped with Ion-Selective Electrodes (Na+, K+, Cl-), ensure slope calibration voltages meet Nernst equation limits specified in Selectra ProS Chemistry Analyzer Operator Manual.

TECHNICAL FAQ

Frequently Asked Questions: Selectra ProS LIS Interfacing

Common biomedical engineering questions regarding connection, worklist setup, and data capture with AILS Cloud LIMS.

Q: What pinouts and network settings are required for the Selectra ProS LIS connection?

The Selectra ProS interfaces using a serial RS-232C Serial DB9 / USB connection with parameters configured to 9600 / 19200 bps, 8-N-1. A dedicated null-modem DB9/DB25 cable with hardware flow control connects to the AILS Gateway hardware edge device or COM-to-IP terminal server.

Q: What is the workflow when loading barcoded tubes onto the Selectra ProS?

The Selectra ProS executes automated positive specimen identification. AILS Gateway matches tube accession numbers against laboratory orders, routing priority STAT orders ahead of routine tests.

Q: How does AILS LIMS process serum indices (Hemolysis, Icterus, Lipemia) from the Selectra ProS?

The Selectra ProS pre-scans specimens and transmits HIL indices alongside chemistry values. AILS LIMS evaluates whether interferent levels exceed analyte-specific thresholds, automatically holding affected tests and alerting lab supervisors.

Q: How does AILS manage Westgard multirule QC validation on the Selectra ProS?

AILS LIMS verifies daily calibration and control runs for Selectra ProS. If QC values drift outside acceptable standard deviation limits, the system interlocks patient reporting until a re-calibration run is approved.

INSTRUMENT DIRECTORY

Related Biochemistry Instrument Guides

Compare protocols, pinouts, and parameter mapping dictionaries for comparable diagnostic equipment.

View All 50+ Analyzers
OFFLINE-RESILIENT GATEWAY

Never Lose a Test Result: Resilient Selectra ProS Middleware

AILS Edge Gateway buffers all data frames locally if your laboratory internet connection drops. Once reconnected, all pending records from the Selectra ProS synchronize automatically to the cloud without human intervention or data loss.