Selectra ProS Bi-Directional LIS Interfacing Guide

Complete engineering specification, HL7/ASTM protocols, pinouts, and parameter mapping for Selectra ProS on AILS Cloud LIMS.

Selectra ProS Analyzer - AILS LIS Interfacing Guide
Manufacturer
Vital Scientific
Protocol
ASTM E1381 / E1394
Interface Mode
Uni-Directional / Bi-Directional
Throughput
140 tests/hour

1. Selectra ProS Interfacing Overview & Diagnostic Workflow

The Selectra ProS is a high-performance clinical chemistry analyzer dedicated to routine photometric assays, enzymatic kinetics, and ion-selective electrode (ISE) electrolytes. Connecting the Selectra ProS to AILS Cloud LIMS establishes a closed-loop analytical cycle—automating test profile assignment, monitoring serum index flags (Hemolysis, Icterus, Lipemia), and preventing reporting errors across high-volume chemistry workloads.

Bi-Directional Host Query & Rack Loading

As sample tubes enter the Selectra ProS reaction carousels, the onboard barcode reader queries AILS Gateway via ASTM E1381 / E1394 to pull ordered test profiles (e.g. Liver Function, Renal Panel, Lipid Profile) in under 300ms.

Serum Index & Auto-Dilution Reflux

AILS automatically reads HIL (Hemolysis, Icterus, Lipemia) interferent flags emitted by the Selectra ProS, warning technologists when sample integrity compromises photometric accuracy and logging auto-dilution rerun factors.

2. Selectra ProS RS-232 Serial Port & Pinout Configuration

Hardware & Socket Configuration

TCP/IP Ethernet (Port 5002) / RS-232 DB9 Serial Port (9600 Baud, 8-N-1).

Protocol: ASTM E1381 / E1394
Interface Mode: Uni-Directional / Bi-Directional
Default Speed: 9600 bps
// RS-232 DB9 NULL-MODEM PINOUT SERIAL COM
[Instrument DB9]                   [AILS Gateway PC]
Pin 2 (TxD / Data Out) ---------> Pin 2 (RxD / Data In)
Pin 3 (RxD / Data In)  <--------- Pin 3 (TxD / Data Out)
Pin 5 (GND / Signal)   <========> Pin 5 (GND / Signal)
Pin 7 (RTS) & Pin 8 (CTS)        [Loopback or None]
Baud: 9600 bps | Data Bits: 8 | Parity: None | Stop: 1
Verified with AILS Gateway Daemon & Real-Time Socket Router.

3. Selectra ProS ASTM E1394 Record Frame

Below is a validated communication frame transmitted between the Selectra ProS and AILS LIS middleware:

<STX>1H|\^&|||Selectra ProS|||||||P|1394-97|20260815141000<CR><ETX>4B<CR><LF>
<STX>2P|1||||Clark^Susan|||F<CR><ETX>2A<CR><LF>
<STX>3O|1|ACC2026081518||^^^BIOCHEM|||||||A<CR><ETX>4C<CR><LF>
<STX>4R|1|^^^GLU|98.5|mg/dL|70.0-100.0|N||F<CR><ETX>1A<CR><LF>
<STX>5R|2|^^^CHOL|188.0|mg/dL|140.0-200.0|N||F<CR><ETX>2B<CR><LF>
<STX>6L|1|N<CR><ETX>0A<CR><LF>

4. Selectra ProS Clinical Parameter LOINC & Reference Range Dictionary

Instrument Code LOINC Code Clinical Parameter Name Standard Units Default Reference Range
GLU 2345-7 Glucose Fasting mg/dL 70.0 - 100.0
CHOL 2093-3 Total Cholesterol mg/dL 140.0 - 200.0

5. Vital Scientific Biomedical Engineering Field Notes & Gotchas

1

Configure Selectra ProS to communicate via TCP port 5002 in Host Settings.

2

Check that reagent barcoded bottles are scanned before launching analytical runs.

3

Ensure AILS Gateway is configured with matching STX/ETX framing.

TECHNICAL FAQ

Frequently Asked Questions: Selectra ProS LIS Interfacing

Common biomedical engineering questions regarding connection, worklist setup, and data capture with AILS Cloud LIMS.

Q: What interface protocols are supported by Selectra ProS?

Selectra ProS supports ASTM E1381/E1394 over TCP/IP socket (Port 5002) and RS-232 serial.

Q: How does Selectra ProS transmit biochemistry test batches?

Results are transmitted in real time as ASTM record frames as each photometric reading completes.

Q: Can Selectra ProS operate with bi-directional worklists?

Yes. Sample barcodes scanned on the rotor are queried against AILS LIMS to fetch test profiles.

Q: How are ISE module parameters mapped on Selectra ProS?

Sodium, Potassium, and Chloride values are mapped to dedicated LOINC codes in AILS LIMS.

Q: What baud rate is used for Selectra ProS serial connections?

Standard baud rate is 9600 bps, 8 data bits, no parity, 1 stop bit.

INSTRUMENT DIRECTORY

Related Biochemistry Instrument Guides

Compare protocols, pinouts, and parameter mapping dictionaries for comparable diagnostic equipment.

View All 50+ Analyzers
PRE-CONFIGURED DRIVER READY

Connect Your Selectra ProS to AILS in Under 15 Minutes

Our specialized biomedical engineering team provides complete plug-and-play drivers, cable diagrams, and bi-directional test verification for the Selectra ProS.