HORIBA / ABX ANALYZER INTERFACING GUIDE

ABX Pentra 60C+ Bi-Directional LIS Interfacing Guide

Complete engineering specification, HL7/ASTM protocols, pinouts, and parameter mapping for ABX Pentra 60C+ on AILS Cloud LIMS.

ABX Pentra 60C+ Analyzer - AILS LIS Interfacing Guide
Manufacturer
Horiba / ABX
Protocol
MS Access ODBC / RS232
Interface Mode
Uni-Directional
Throughput
60 samples/hour

1. Instrument Overview & Interfacing Architecture

The ABX Pentra 60C+ is a high-performance hematology instrument designed for diagnostic laboratories and hospital health networks. Integrating this analyzer directly with AILS Cloud LIMS eliminates manual transcription delays, prevents typographical patient identification errors, and automatically enforces Westgard delta check algorithms.

Bi-Directional Order Push

AILS automatically generates worklist queries (QRY^Q02 / ASTM orders) mapped to patient accession barcodes and pushes test instructions directly to the instrument sample tray.

Real-Time Result Ingestion

Raw OBX / Result records are parsed instantly within 200 milliseconds, flagged for abnormal reference ranges, and linked to pathologist digital signature queues.

2. Hardware Connection & Communication Settings

// Physical & Socket Configuration
MS Access ODBC Database Sync Agent / RS232 Serial Port (9600 Baud, 8-N-1).
// Protocol: MS Access ODBC / RS232 | Connection Type: ODBC DB Sync & Serial COM

3. Sample Message Wire Frame

Below is a real-world communication frame transmitted between the ABX Pentra 60C+ and AILS LIS middleware:

[ODBC_RESULT_POLL_RECORD]
SampleID: ACC2026081524
PatientName: Davis, Michael
RunTimestamp: 2026-08-15 14:20:12
WBC: 7.15 (10^3/uL)
RBC: 4.85 (10^6/uL)
HGB: 15.0 (g/dL)
PLT: 220 (10^3/uL)
Status: Validated

4. Parameter & LOINC Standard Code Mapping

Instrument Code LOINC Code Clinical Parameter Name Standard Units Default Reference Range
WBC 6690-2 White Blood Cells 10*3/uL 4.0 - 11.0
RBC 789-8 Red Blood Cells 10*6/uL 4.5 - 5.9
HGB 718-7 Hemoglobin g/dL 13.5 - 17.5

5. Biomedical Engineering Tips & Common Gotchas

1

Ensure Horiba local data export utility is actively flushing completed runs to intermediate DB.

2

AILS background gateway polls table 'RESULTS' every 2000ms with transaction locking.

3

Verify that 5-part differential DHSS matrix data is exported.

TECHNICAL FAQ

Frequently Asked Questions: ABX Pentra 60C+ Interfacing

Common biomedical engineering questions regarding connection, worklist setup, and data capture with AILS Cloud LIMS.

Q: How does ODBC integration work for Horiba Pentra 60C+?

AILS Gateway connects directly to the local MS Access (.mdb) database updated by the Pentra workstation, polling new completed runs automatically.

Q: Can Pentra 60C+ connect via RS-232 serial cable instead of ODBC?

Yes. Pentra 60C+ also supports direct serial COM output at 9600 baud using the standard ABX protocol.

Q: How frequently does AILS Gateway poll the Pentra 60C+ database?

The default polling interval is 2000 milliseconds, ensuring real-time result ingestion with zero server lag.

Q: Are 5-part differential parameters captured via ODBC?

Yes. Neutrophils, Lymphocytes, Monocytes, Eosinophils, and Basophils (# and %) are extracted from table records.

Q: What happens if the local computer restarts?

AILS Gateway runs as a resilient Windows service with automatic startup and crash recovery.

PRE-CONFIGURED DRIVER READY

Connect Your ABX Pentra 60C+ to AILS in Under 15 Minutes

Our specialized biomedical engineering team provides complete plug-and-play drivers, cable diagrams, and bi-directional test verification for the ABX Pentra 60C+.