HORIBA / ABX ANALYZER INTERFACING GUIDE

ABX Micros 60 Bi-Directional LIS Interfacing Guide

Complete engineering specification, HL7/ASTM protocols, pinouts, and parameter mapping for ABX Micros 60 on AILS Cloud LIMS.

ABX Micros 60 Analyzer - AILS LIS Interfacing Guide
Manufacturer
Horiba / ABX
Protocol
Serial Stream (RS-232)
Interface Mode
Uni-Directional
Throughput
60 samples/hour

1. Instrument Overview & Interfacing Architecture

The ABX Micros 60 is a high-performance hematology instrument designed for diagnostic laboratories and hospital health networks. Integrating this analyzer directly with AILS Cloud LIMS eliminates manual transcription delays, prevents typographical patient identification errors, and automatically enforces Westgard delta check algorithms.

Bi-Directional Order Push

AILS automatically generates worklist queries (QRY^Q02 / ASTM orders) mapped to patient accession barcodes and pushes test instructions directly to the instrument sample tray.

Real-Time Result Ingestion

Raw OBX / Result records are parsed instantly within 200 milliseconds, flagged for abnormal reference ranges, and linked to pathologist digital signature queues.

2. Hardware Connection & Communication Settings

// Physical & Socket Configuration
DB9 Female Serial Cable (COM2). Baud: 9600 bps, 8 Data Bits, Parity None, 1 Stop Bit.
// Protocol: Serial Stream (RS-232) | Connection Type: RS232 Serial (COM2 / 9600)

3. Sample Message Wire Frame

Below is a real-world communication frame transmitted between the ABX Micros 60 and AILS LIS middleware:

[ABX_MICROS_60_STREAM]
SAMPLE_ID: ACC2026081526
DATE: 2026/08/15  14:30
WBC: 7.2 10^3/uL
RBC: 4.80 10^6/uL
HGB: 14.6 g/dL
HCT: 43.0 %
PLT: 235 10^3/uL
LYM%: 29.0 %
MON%: 6.0 %
GRA%: 65.0 %

4. Parameter & LOINC Standard Code Mapping

Instrument Code LOINC Code Clinical Parameter Name Standard Units Default Reference Range
WBC 6690-2 White Blood Cells 10*3/uL 4.0 - 11.0
HGB 718-7 Hemoglobin g/dL 13.5 - 17.5
PLT 777-3 Platelets 10*3/uL 150 - 450

5. Biomedical Engineering Tips & Common Gotchas

1

In Micros 60 rear settings, ensure RS-232 output mode is set to Host/Computer.

2

AILS Gateway captures the serial string automatically upon cycle completion.

3

Verify COM port assignment in Windows Device Manager.

TECHNICAL FAQ

Frequently Asked Questions: ABX Micros 60 Interfacing

Common biomedical engineering questions regarding connection, worklist setup, and data capture with AILS Cloud LIMS.

Q: What communication interface is used by Horiba ABX Micros 60?

ABX Micros 60 outputs an 18-parameter structured ASCII text record over its RS-232 serial port at 9600 baud.

Q: What cable is required for ABX Micros 60?

A standard DB9 Null Modem serial cable connected to the analyzer's serial port.

Q: Does AILS capture 3-part differential results from Micros 60?

Yes. Lymphocyte, Monocyte, and Granulocyte counts (# and %) are parsed into LOINC codes.

Q: Can USB-to-Serial adapters be used with Micros 60?

Yes. High-grade FTDI USB-to-RS232 adapters are supported for modern PCs.

Q: How does AILS handle micro-sampling modes (e.g. pediatric capillary blood)?

Capillary and pre-diluted sample flags are recorded in the patient report.