GETEIN ANALYZER INTERFACING GUIDE

Getein 1100 / 1600 FIA Analyzer Bi-Directional LIS Interfacing Guide

Complete engineering specification, HL7/ASTM protocols, pinouts, and parameter mapping for Getein 1100 / 1600 FIA Analyzer on AILS Cloud LIMS.

Getein 1100 / 1600 FIA Analyzer Analyzer - AILS LIS Interfacing Guide
Manufacturer
Getein
Protocol
RS-232 Serial / HL7 Protocol
Interface Mode
Uni-Directional
Throughput
Rapid Test Strips Ingest

1. Getein 1100 / 1600 FIA Analyzer Interfacing Overview & Diagnostic Workflow

The Getein 1100 / 1600 FIA Analyzer is a precision diagnostic analyzer designed for clinical laboratory testing. Connecting the Getein 1100 / 1600 FIA Analyzer directly to AILS Cloud LIMS eliminates manual data entry, prevents typographical identification errors, and delivers verified results directly to ordering physicians and hospital EMRs.

Automated Result Stream Capture

AILS Gateway continuously monitors the Getein 1100 / 1600 FIA Analyzer connection, capturing transmitted test observations within 200 milliseconds of analytical completion.

Quality Control & Reference Range Interlock

Results are checked against age- and gender-adjusted reference ranges, delta check baselines, and quality control (QC) Levey-Jennings rules before digital sign-off.

2. Getein 1100 / 1600 FIA Analyzer TCP/IP Socket & Network Configuration

Hardware & Socket Configuration

RS-232 DB9 Serial Port (COM6 / 9600 Baud) or TCP/IP Socket. HL7 v2.3.1 or Text Stream.

Protocol: RS-232 Serial / HL7 Protocol
Interface Mode: Uni-Directional
Default Speed: 9600 bps, 8-N-1
// RS-232 DB9 NULL-MODEM PINOUT SERIAL COM
[Instrument DB9]                   [AILS Gateway PC]
Pin 2 (TxD / Data Out) ---------> Pin 2 (RxD / Data In)
Pin 3 (RxD / Data In)  <--------- Pin 3 (TxD / Data Out)
Pin 5 (GND / Signal)   <========> Pin 5 (GND / Signal)
Pin 7 (RTS) & Pin 8 (CTS)        [Loopback or None]
Baud: 9600 bps | Data Bits: 8 | Parity: None | Stop: 1
Verified with AILS Gateway Daemon & Real-Time Socket Router.

3. Getein 1100 / 1600 FIA Analyzer HL7 v2.x Message Stream

Below is a validated communication frame transmitted between the Getein 1100 / 1600 FIA Analyzer and AILS LIS middleware:

[GETEIN_1100_SERIAL_RECORD]
SampleID: ACC2026081584
TestTime: 2026-08-15 18:10:00
Item: NT-proBNP
Result: 850.00 pg/mL
RefRange: 0.00 - 125.00 pg/mL
Flag: H
Item: cTnI
Result: 0.25 ng/mL
RefRange: 0.00 - 0.04 ng/mL
Flag: H

4. Getein 1100 / 1600 FIA Analyzer Clinical Parameter LOINC & Reference Range Dictionary

Instrument Code LOINC Code Clinical Parameter Name Standard Units Default Reference Range
NT-proBNP 33762-6 NT-proBNP Cardiac Marker pg/mL 0.00 - 125.00
cTnI 10839-9 Cardiac Troponin I ng/mL 0.00 - 0.04
hs-CRP 30522-7 hs-CRP + Routine CRP mg/L 0.00 - 3.00
D-Dimer 48065-7 D-Dimer (FEU) mg/L 0.00 - 0.50
PCT 33959-8 Procalcitonin ng/mL 0.00 - 0.50

5. Getein Biomedical Engineering Field Notes & Gotchas

1

In Getein 1100 setup, configure Data Transmission to LIS Mode with Baud Rate 9600.

2

Ensure SD card calibration curve chip is inserted before running new reagent test lots.

3

AILS Gateway captures both single-strip and multi-parameter combo test cartridge records.

TECHNICAL FAQ

Frequently Asked Questions: Getein 1100 / 1600 FIA Analyzer LIS Interfacing

Common biomedical engineering questions regarding connection, worklist setup, and data capture with AILS Cloud LIMS.

Q: What interface protocols are supported by Getein 1100 and Getein 1600?

Getein 1100 uses RS-232 serial stream communication at 9600 baud. Getein 1600 supports both RS-232 serial and HL7 v2.3.1 over TCP/IP socket.

Q: How does AILS handle multi-parameter combo test cards (e.g. cTnI/CK-MB/Myo)?

AILS parses multi-analyte cartridges into individual sub-test results under a consolidated cardiac emergency panel.

Q: What serial cable is required for Getein 1100?

A standard DB9 Female Null Modem serial cable connected to COM6 on the gateway PC.

Q: How does AILS handle urgent STAT panic values from emergency room Getein readers?

Elevated cardiac biomarkers trigger visual emergency banners on central laboratory dashboards and send automated WhatsApp/SMS alerts to clinicians.

Q: Can Getein results be auto-verified in AILS LIMS?

Yes. If all quantitative POC parameters fall within normal physiological limits, AILS can automatically verify and digitally sign the report.

INSTRUMENT DIRECTORY

Related Point-of-Care (POC) Instrument Guides

Compare protocols, pinouts, and parameter mapping dictionaries for comparable diagnostic equipment.

View All 50+ Analyzers
PRE-CONFIGURED DRIVER READY

Connect Your Getein 1100 / 1600 FIA Analyzer to AILS in Under 15 Minutes

Our specialized biomedical engineering team provides complete plug-and-play drivers, cable diagrams, and bi-directional test verification for the Getein 1100 / 1600 FIA Analyzer.